MAX ILLUMS ACTIVE
This message will be displayed if you attempt to activate more than
20 illuminants at a time. Due to limits in the memory system, a
maximum of 20 illuminant/observer combinations may be active at
a time.
MAX FORMULAS ACTIVE
This message will be displayed if you attempt to activate too many
illuminants at a time. Due to limits in the memory system, a
maximum number of illuminant/observer combinations may be
active at a time.
CANNOT REMOVE ALL
This message will be displayed if you attempt to disable all
illuminants or all formulas. These types of data require that at least
one of each type be enabled.
INVALID DATE
This message indicates an invalid date has been entered as the
system date. Modify the date to reflect a valid date.
CHECKER TILE VALUE OUT OF RANGE
This message indicates the value entered as the Illuminant
Checker’s CIE Whiteness value is out of range. The value must be
greater than zero and less than 999.99. Check the value and enter it
again.
UVCAL STEP OUT OF RANGE
This message indicates the UVCAL step position is invalid. The
minimum step number for the UVCAL position is four. The
maximum step number depends on the motor used to control the
UV filter position. If entering the step position manually, input a
different number. If performing an operation which determines the
UVCAL step position automatically, this may indicate a possible
instrument malfunction and your Datacolor International Service
representative should be contacted for further details.