Basic Operation: Absorbance
Best Practices Absorbance Measurements
Clean both sample measurement surfaces prior to making the Blank
measurement.
Use dH
2
O, not detergents or alcohol, for routine sample measurement surface
cleaning.
Use a dry lab wipe to remove measurement liquids from both the top and
bottom surfaces immediately after each measurement.
Use 1 μL samples for routine microvolume measurements.
Note: Select the Short Path mode for high concentration 0.5 μL samples.
Use a fresh aliquot for each measurement.
Use a fresh tip to deliver each sample aliquot.
Avoid introducing bubbles when pipetting samples onto the measurement
surfaces.
Ensure sample concentrations fall within absorbance limits of the instrument
for accurate results.
Use the same buffer a sample is suspended in for the Blank measurements.
Avoid using buffers such as RIPA buffers that contain components with strong
absorbances at the wavelength of interest.
Measure a fresh aliquot of the buffer using the Measure button and ensure the
spectrum is relatively flat along the baseline before measuring samples.
For cuvette mode measurements, use cuvettes with 8.5 mm Z heights. Select
the Cuvette option using the drop-down menu next to the Measure button and
ensure the cuvette is inserted in the proper orientation.$