Fluorescence Basic Operation
Best Practices
Ensure all sample concentrations in the thin walled assay tubes fall within the
limits of the assay or reagent kit for accurate results.
Use 200 μL volumes in a clear thin walled PCR tube.
Lower the cover before tapping the measure button.
Avoid introducing air bubbles into the sample solution when mixing samples.
Follow assay reagent manufacturers' recommendations regarding
temperature, incubation time, and protection from ambient light.
Follow assay reagent manufacturers' recommendations regarding suggested
standard curve concentrations and data analysis.
Ensure all samples and standards are treated identically in terms of incubation
times and temperature.
Minimize assay tube and solution temperature fluctuations as these may
impact the accuracy of the assay. Samples at higher temperatures will
generally have lower fluorescence.
Do not label the side of an assay tube as this could interfere with the sample
measurement.