Generating a Standard Curve
1. Use the drop-down menu on the Std Curve screen to select New Curve or to
import a saved standard curve or standard curve concentration list.
2. If the New Curve option is selected, tap the Enter Standards button to
enable a pop-up box to enter a list of standard concentrations for the current
measurement session.
• The table enables up to 7 standard concentrations in addition to the
predefined 0.00.
• The 0.00 concentration is used for the diluent (no analyte) plus the assay
reagent. Although not recommended, the 0.00 concentration may be
edited.
3. Measure at least one replicate of each standard concentration by highlighting
the standard concentration row and then tapping the Measure button.
• Up to 5 replicates may be measured for each concentration. Use fresh
aliquots for each measurement when using the microvolume mode.
• The software will advance after each measurement until the fifth replicate
has been measured. If additional measurements are made, the fifth
position will be overwritten.
• A minimum of two standards must be measured before the Measure
button is active on the Run screen.
• Select a standard row or individual replicate then tap the Delete icon to
display a drop-down list of delete options.
• Use the Undo icon to restore the most recent edit or delete action.