EasyManuals Logo

DeNovix DS-11 Series User Manual

DeNovix DS-11 Series
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
21. Troubleshooting
Quick Help
Most issues concerning accuracy, reproducibility, negative spectra and low
nucleic acid purity ratios are sample or technique related and are resolved by
following the suggestions below:
Establish a new Blank using the appropriate buffer.
Ensure that the sample isolation procedure is optimized and that samples are
purified when required prior to making absorbance measurements.
Ensure all solutions are homogenous and well-mixed prior to sampling.
Ensure sample concentrations fall within the absorbance limits of the
instrument. In the case of fluorescence assays, ensure sample concentrations
fall within the reagent limits as described by the assay manufacturer.
Microvolume Mode
Ensure both top and bottom microvolume measurement surfaces are clean
prior to making the Blank measurement.
Always remove sample solutions from both the top and bottom measurement
surfaces using dry, lint-free wipe immediately after each measurement is
complete.
Use calibrated pipettors and properly fitting tips to ensure a full 1 μL aliquot is
delivered to the sample measurement surface. Protein samples sometimes
wick up on the outside of the tip and may not be properly dispensed.
Use a clean pipette tip and a new aliquot of sample for each microvolume
measurement. $
90

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DeNovix DS-11 Series and is the answer not in the manual?

DeNovix DS-11 Series Specifications

General IconGeneral
Wavelength Range190 - 840 nm
Minimum Sample Size0.5 µL
Sample Volume0.5 - 2 µL
Absorbance Precision0.002 AU (1 cm equivalent)
Operating SystemAndroid
Display7-inch touchscreen
Power Requirements100 - 240 V, 50/60 Hz
Pathlength0.5 mm (auto-ranging)