Cuvette Mode
Use cuvettes with Z heights of 8.5 mm.
Use quartz or UV transparent cuvettes for UV range measurements.
Ensure cuvettes are clean prior to use.
Fill cuvettes to manufacturer's specifications.
Confirm the mode selection has been changed on the Run screen to the
proper cuvette pathlength prior to a measurement.
Ensure cuvettes are properly inserted following the etched arrow light path.
Refer to a reagent kit manufacturer for troubleshooting colorimetric assays.
Fluorescence Mode
Use 200 μL volumes in a clear thin walled PCR tube
Avoid introducing air bubbles into the sample solution when mixing samples.
Ensure all sample concentrations in the thin walled assay tubes fall within the
limits of the assay or reagent kit for accurate results.
Follow assay reagent manufacturers' recommendations regarding
temperature, incubation time, and protection from ambient light.
Follow assay reagent manufacturers' recommendations regarding suggested
standard curve concentrations and data analysis.
Confirm that the correct LED excitation source was used for all user-defined
assays and methods.
Confirm that standard concentrations and dilutions were performed correctly.
Confirm that the correct concentration units for the standard curve and the
unknown samples were used to calculate the stock concentrations.
91

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DeNovix DS-11 Series and is the answer not in the manual?

DeNovix DS-11 Series Specifications

General IconGeneral
Wavelength Range190 - 840 nm
Minimum Sample Size0.5 µL
Sample Volume0.5 - 2 µL
Absorbance Precision0.002 AU (1 cm equivalent)
Operating SystemAndroid
Display7-inch touchscreen
Power Requirements100 - 240 V, 50/60 Hz
Pathlength0.5 mm (auto-ranging)