Custom Methods: Absorbance
Custom Formula Methods
In addition to the primary analysis wavelength, the Formula Methods app enables
the user to include up to 4 additional results for each measurement. The result
fields may be used to monitor additional specific wavelengths or have the
software perform routine calculations based upon absorbance values measured
at specific wavelengths.
Result Name: Enter a name for each custom formula.
Custom Formulas: Tap on the custom formulas input field to display a pop-up
dialog.The following functions are available:
Advanced: When checked, this feature allows the user to designate that the
method uses a Savitzky–Golay filter instead of the default boxcar method to
smooth data. $
51
Function
Description
Numerical Operations
+ add - subtract * multiply ÷ divide
A(x)
Absorbance at wavelength x
Log(x)
Log of x
ln(x)
Natural Log of x
^
Exponent
Example 4^3 equals 4x4X4
Square Root
e
Scientific notation
Example 2e3 equals 2 x 10
3
or 2000
V1, V2
Variables that are entered on the Run screen
prior to each measurement.