Nucleic Acids: Absorbance
dsDNA, RNA and ssDNA
The basic nucleic acid apps are optimized for the factor used to calculate sample
concentrations.
Screen Features
Sample Type: The sample type is used to determine which ng-cm/μL factor is
used to calculate the concentration.
The sample type is fixed for both the dsDNA and RNA apps.
A Factor Input option along with the default ssDNA sample type is
available from the selector field for the ssDNA app.
Factor: The default wavelength-dependent factors used for each app are as
follows: dsDNA: 50 ng-cm/μL RNA: 40 ng-cm/μL ssDNA: 33 ng-cm/μL
Oligo factors derived from the Calculator app may be copied and pasted
directly into this field.
Concentration Results: Concentration results are reported in units of ng/μL.
Units may be changed using the Run screen top action bar Overflow
menu. The units selected at the time of the measurement is what is saved
to the database.
A260: Reports the baseline-corrected 260 nm absorbance value.
260/230: Displays the ratio of the 260 nm and 230 nm absorbance values.
260/280: Displays the ratio of the 260 nm and 280 nm absorbance values.
Note: The DNA and RNA apps do not distinguish between the nucleic acid
species within a sample. It is important to optimize the isolation protocol to
ensure accurate quantitation.$
21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DeNovix DS-11 Series and is the answer not in the manual?

DeNovix DS-11 Series Specifications

General IconGeneral
Wavelength Range190 - 840 nm
Minimum Sample Size0.5 µL
Sample Volume0.5 - 2 µL
Absorbance Precision0.002 AU (1 cm equivalent)
Operating SystemAndroid
Display7-inch touchscreen
Power Requirements100 - 240 V, 50/60 Hz
Pathlength0.5 mm (auto-ranging)