EasyManua.ls Logo

DeNovix DS-11 Series - 19. Maintenance; Cleaning; Guidelines

DeNovix DS-11 Series
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19. Maintenance
Cleaning
Guidelines
Use only a dry, soft, lint-free cloth to clean the front screen.
Do not use a spray bottle to apply water or any other solutions onto any
surface of the instrument as the liquid may damage internal components.
The use of detergents or isopropyl alcohol as cleaning solvents is not
recommended as they may temporarily alter the hydrophobic nature of the
sample surfaces.
Routine
1. Pipette 2 μL of dH
2
0 onto the bottom sample surface. Lower
the upper arm. Wait 1-2 minutes.
2. Wipe away the water from both the upper and lower sample
surfaces with a dry, lint-free lab wipe.
Between
Measurements
Use a dry, lint-free lab wipe to remove samples from both the
lower and upper sample surfaces.
Additional
Cleaning
Use the following procedure to remove sticky solutions (e.g.
proteins) that may have dried down on the sample surfaces.
1. Pipette 2 μL of 0.5 M HCl onto the bottom sample surface.
Lower the upper arm. Wait 2-3 minutes.
2. Wipe away the HCl from both the upper and lower sample
surfaces with a dry, lint-free lab wipe.
3. Repeat steps 1-2 with 2 μL of dH
2
0.
83