OD 600
Unique Features
A600: The OD 600 is a measure of the light scattered by the cell suspension
solution. Reported values may differ for the same sample measured on
different spectrophotometer systems as values are dependent on both the cell
type and the optical configuration of the spectrophotometer.
o It is recommended that an optimal harvest density and the
linear range of growth curves be empirically determined for
each microbial cell type when using a new spectrophotometer.
Cell Number Conversion Factor: The cell number conversion factor is used
to convert the 600 nm value into the number of cells per mL for the sample by
multiplying the absorbance value by the factor. Cell numbers are reported in
terms of 1 * 10
8
cells. The factor must be entered into the appropriate field
prior to a measurement.
Note: The cuvette mode selection (i.e the path length) must be taken into
account when determining target OD 600 values or entering cell conversion
factors. When making microvolume measurements, the OD 600 values
reported and used for cell number calculations are 10 mm equivalent values.
Baseline Correction
As microbial cell culture OD 600 values are measurements of light scattering, the
DS-11 software does not use a bichromatic baseline normalization unless
specified as a user selection.
The baseline correction screen is accessed from the Overflow icon. Once a
baseline correction nm is selected, it is applied to all subsequent measurements
within the same measurement session.
44

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DeNovix DS-11 Series and is the answer not in the manual?

DeNovix DS-11 Series Specifications

General IconGeneral
Wavelength Range190 - 840 nm
Minimum Sample Size0.5 µL
Sample Volume0.5 - 2 µL
Absorbance Precision0.002 AU (1 cm equivalent)
Operating SystemAndroid
Display7-inch touchscreen
Power Requirements100 - 240 V, 50/60 Hz
Pathlength0.5 mm (auto-ranging)