Kinetics: Absorbance
Run Screen Features
Method: Select the method of choice using the drop-down menu. Use the List
Add/Edit button to the right of the drop-down to create or edit methods. All
methods are available to all accounts. However, only the account holder or an
administrator may edit or delete a method.
Graph Type Flip Button: Graph types may be switched during the run.
Spectra: Absorbance vs wavelength
Trend: Absorbance vs elapsed time for analysis wavelength(s)
Heater Control: Tap the On/Off button to control the heater. The heater may
overshoot the method set point before reaching a equilibrium + 0.5ºC of the set
point.
Timer: This feature is used to monitor the sample solution warm-up time. The
timer is used in a count-up only mode but is not used to track the
measurement time.
Stop Button: A Stop button replaces the Measure button on the Run screen
during the Kinetics run. Tapping this button will end the measurement and save
all data accumulated up to that point. The Measure button will be available
when the run is finished.
Results Table: A scrollable table with the absorbance values at the specified
wavelength(s) per time is displayed on the Run screen.
Heater Temperature
The temperature display box is color-coded for quick reference:
White: Heater is off.
Yellow: Heater is on, heater is not at method set point.
Orange: Heater is on, cuvette holder block has reached the method set
point within + 0.5ºC.
47