How to fix E06 error on DeVilbiss DV63 Standard Plus Medical Equipment?
- JJustin HaynesAug 26, 2025
If your DeVilbiss Medical Equipment displays an E06 error code and the blower doesn't start, first inspect the air intake filter, intake foam components, and internal connections. Replace any damaged foam pieces. Next, check for any signs of water damage, especially on the blower assembly, pressure sensor, or flow sensor, and replace any damaged components. Ensure the blower halves are properly snapped together without any visible gaps. Then, recalibrate the pressure and flow sensors using the Manual Calibration Procedure. Finally, test with a known good blower. If the good blower runs, replace the original blower and perform a 6-hour burn-in test with sound testing. If the good blower doesn't run, replace the control board.