EasyManua.ls Logo

Dexcom G5 Mobile - Page 107

Dexcom G5 Mobile
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
105Calibration
Warning
Do: Calibrate at least once every 12 hours.
Why: Calibrating less often than every 12 hours might cause inaccurate sensor glucose
readings.
Consequences: Missing a severe low or high blood glucose event or making a
treatment decision that results in injury.
Precaution
Do: Look at trend arrows before calibrating. Trend arrows help you determine if you can
calibrate now or should wait.
Don’t: Never calibrate if your BG is changing at a significant rate, typically more than 2
mg/dL per minute.
Never calibrate when you see:
• A single arrow, pointing up
– Rising 2-3 mg/dL each minute
• Two Arrows pointing up
– Rising more than 3 mg/dL each minute
• Single arrow pointing down
– Falling 2-3 mg/dL each minute
• Two arrows pointing down
– Falling more than 3 mg/dL each minute
Why: Calibrating during significant rise/fall of BG may affect accuracy of sensor
glucose readings.
Consequences: Missing a severe low or high blood glucose event or making a
treatment decision that results in injury.

Table of Contents