19Indications for Use and Safety Statement
Contraindications
MRI/CT/ Diathermy
Remove the Dexcom G5 sensor, transmitter, and receiver before Magnetic Resonance
Imaging (MRI), Computed Tomography (CT) scan, or high-frequency electrical heat
(diathermy) treatment.
The Dexcom G5 has not been tested during MRI or CT scans or with diathermy treatment.
The magnetic fields and heat could damage the components of the Dexcom G5, which may
cause it to display inaccurate blood glucose readings or may prevent alerts.
Medications
Taking medications with acetaminophen while wearing the Dexcom G5 may inaccurately
raise the glucose readings generated by the Dexcom G5. The level of inaccuracy depends
on the amount of acetaminophen active in your body and is different for each person. Do not
rely on continuous glucose monitoring (CGM) data produced by the Dexcom G5 if you have
recently taken acetaminophen.
Warnings
Sensor Fractures
Do not ignore sensor fractures. Sensors may fracture or detach from the sensor pod on
rare occasions. If a sensor breaks and no portion of it is visible above the skin, do not
attempt to remove it. Seek professional medical help if you have symptoms of infection or
inflammationredness, swelling, or painat the insertion site. If you experience a broken
sensor, please report this to our Technical Support department at 1.888.738.3646 (toll free)
or 1.858.200.0200.
Do Not Use Damaged Goods
If the Dexcom G5 receiver or Dexcom G5 transmitter is damaged or cracked, do not use it.
This could create an electrical safety hazard causing possible electrical shocks resulting in
injury. In addition, a damaged or cracked Dexcom G5 receiver or Dexcom G5 transmitter may
cause the Dexcom G5 System not to function properly.
Choking
Do not allow young children to hold the sensor or transmitter without adult supervision. The
sensor and transmitter include small parts that may pose a choking hazard.