Dexcom G5 Mobile System User Guide
307Technical Information
Table 5-C. Hypoglycemic Alert and Detection Rate Evaluation in
Reference to SMBG 30 Minutes Before and After (Pediatric, Ages 2-5
Years)
Hypoglycemic
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hypoglycemia
Detection Rate
Hypoglycemia
Missed
Detection Rate
55
Original 3% 97% 57% 43%
Software 505 25% 75% 100% 0%
60
Original 11% 89% 62% 38%
Software 505 20% 80% 100% 0%
70
Original 29% 71% 77% 23%
Software 505 20% 80% 100% 0%
80
Original 35% 65% 85% 15%
Software 505 61% 39% 100% 0%
90
Original 51% 49% 89% 11%
Software 505 78% 22% 100% 0%
100
Original 64% 36% 91% 9%
Software 505 82% 18% 100% 0%
1
Both sets of study data are presented and are labeled as Original (SW10050) or Software 505
(SW10505).
Low and High Glucose Alerts
The ability of the System to detect high and low glucose levels is assessed by comparing System
results to YSI results at low and high blood glucose levels and determining if the alert may have
sounded. The System and YSI values were compared by pairing the System reading that occurred
immediately after the YSI value was collected. We suggest that you ask your doctor what alert settings
would be best for you.
The Low Glucose Alert
Estimates of how well the adjustable Low Glucose Alert performs are presented in Tables 5-A, 5-B and
5-C. Table 5-A represents the hypoglycemic alert evaluation within 15 minutes of the YSI value in the
adult studies. Table 5-B represents the alert evaluation within 15 minutes of the YSI value for a subset
of the pediatric populationsubjects age 6 to 17 years who had YSI measurements every 15 minutes.