EasyManua.ls Logo

Dexcom G5 Mobile - Page 312

Dexcom G5 Mobile
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
310Technical Information
Table 6-B. Hyperglycemic Alert and Detection Rate Evaluation in
Reference to YSI 15 Minutes Before and After (Pediatric, Ages 6-17 Years)
Hyperglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hyperglycemia
Detection Rate
Hyperglycemia
Missed
Detection Rate
120
Original 91% 9% 98% 2%
Software 505 98% 2% 99% 1%
140
Original 87% 13% 99% 1%
Software 505 97% 3% 98% 2%
180
Original 75% 25% 99% 1%
Software 505 94% 6% 98% 2%
200
Original 71% 29% 98% 2%
Software 505 94% 6% 97% 3%
220
Original 67% 33% 97% 3%
Software 505 93% 7% 96% 4%
240
Original 62% 38% 96% 4%
Software 505 88% 12% 94% 6%
300
Original 43% 57% 93% 7%
Software 505 69% 31% 8 4% 16%
1
Both sets of study data are presented and are labeled as Original (SW10050) or Software 505
(SW10505).

Table of Contents

Other manuals for Dexcom G5 Mobile

Related product manuals