Dexcom G5 Mobile System User Guide
313Technical Information
Table 7-A. Percentage of System Readings
1
Within YSI Values With Data
Stratified in 2-Hour Increments After Calibration (Adult)
Time From
Calibration
Study
2
Number
of Paired
CGM-YSI
Percent
Within
15/15%
YSI
Percent
Within
20/20%
YSI
Percent
Within
30/30%
YSI
Percent
Greater
Than
40/40%
YSI
0-2 hours
Original 1929 78% 88% 96% 2%
Software 505 469 93% 97% 99% 0%
2-4 hours
Original 1516 69% 81% 91% 4%
Software 505 389 90% 97% 99% 0%
4-6 hours
Original 1547 69% 79% 91% 5%
Software 505 383 85% 91% 97% 2%
6-8 hours
Original 1520 68% 79% 92% 3%
Software 505 380 79% 90% 97% 2%
8-10 hours
Original 1555 71% 82% 92% 4%
Software 505 347 83% 92% 98% 0%
10-12 hours
Original 1068 65% 77% 91% 4%
Software 505 295 80% 90% 98% 0%
12-14 hours
Original 17 65% 76% 82% 12%
Software 505 0 -- -- -- --
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software 505
(SW10505).