EasyManua.ls Logo

Dexcom G5 Mobile - Page 327

Dexcom G5 Mobile
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dexcom G5 Mobile System User Guide
325Technical Information
Table 12-A. CGM System Difference to SMBG Within CGM Glucose Ranges
(Adult)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-SMBG
Mean
Percent
Difference
Median
Percent
Difference
Mean
Absolute
Percent
Difference
Median
Absolute
Percent
Difference
Overall
Original 7508 -0.4% -1.4% 14.0% 11.0%
Software 505 2992 -2.6% -2.7% 11.3% 8.6%
*40-60
Original 731 -9.3 -8.0 11.7 8.0
Software 505 221 -10.3 -6.0 13.0 8.0
*61-80
Original 968 -1.0 1.0 10.7 8.0
Software 505 336 -4.0 -2.0 10.1 7.0
81-180
Original 3141 1.4% 0.0% 14.2% 11.0%
Software 505 1362 -2.6% -3.1% 11.4% 8.9%
181-300
Original 1960 -0.7% -2.8% 13.0% 10.3%
Software 505 826 -1.4% -2.0% 9.5% 7.4%
301-350
Original 450 -0.7% -2.6% 10.5% 8.6%
Software 505 161 -0.0% 0.0% 8.3% 6.0%
351-400
Original 258 5.0% 3.0% 11.9% 8.6%
Software 505 86 3.9% 3.2% 8.1% 6.7%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or Software 505
(SW10505).
*
For CGM ≤ 80 mg/dL, the differences in mg/dL are included instead of percent differences (%).