82 Appendix G: Technical Information
Dexcom G6
®
System • Using Your G6
G.1 Product Specifications
WARNING: Use of accessories, cables, adapters, and chargers other than those
specified or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm to
any part of the G6 CGM system including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
The device behaves normally while charging, however, do not hold the receiver while
charging for over a minute as the device can get warm to the touch.
No cleaning methods are recommended or tested for the receiver. Only wipe with a
clean, dry cloth.
Sensor Product Specifications
Glucose Range 2.222.2 mmol/L
Sensor Useful Life Up to 10 days
Storage and Transport
Conditions
Temperature: 2°C30°C
Store sensors in a cool, dry place
Sterilization Sterile by radiation
Transmitter and Receiver Product Specifications
Model G6 Transmitter Dexcom Receiver
Memory Storage
30 days of glucose data
7 days of tech support data
Electrical Safety
Class
Internally Powered Internally Powered
Battery Longevity
(Typical)
3 months 2 days