83Appendix G: Technical Information 83
Dexcom G6
®
System • Using Your G6
Transmitter and Receiver Product Specifications
Model G6 Transmitter Dexcom Receiver
Battery Charging
Time
Non-rechargeable 3 hours
Operational
Conditions
Ambient temperature is
10°C42°C
Humidity: 10%95% RH
Temperature: 0°C45°C
Humidity: 15%95% RH
Storage and
Transport
Conditions
Temperature: 0°C45°C
Humidity: 10%95% RH
Temperature: 0°C40°C
Humidity: 10%95%RH
Operating
Altitude
-396 meters to 4,206 meters -365 meters to 4114 meters
Ingress
Protection
IP28: Protection against
insertion of large objects and
immersion in water for up to
2.4 meters for 24 hours
IP22: Protection against
insertion of large objects
and vertically falling water
drops
Protection
Against Electrical
Shock
Type BF applied part N/A
Alarm Audible
Output
N/A 50 dB
SPL
at 1 meter
TX/RX
Frequencies
2.402–2.480 GHz
Bandwidth 1.07 MHz 1.39 MHz
Maximum Output
Power
1.0 mW EIRP 2.4 mW EIRP
Modulation Gaussian Frequency-Shift Keying
Data Rate 1 Mbps
Data
Communication
Range
6 meters