84 Appendix G: Technical Information
Dexcom G6
®
System • Using Your G6
USB Charging/Download Cable* Specifications
Input/Output 5 V DC, 1A
Type USB A to USB micro B
Length 0.91 meters
Power Supply/Charger Specifications
Class II
Input
AC Input 100240 Vac, 50/60Hz, 0.2A, 0.2A rms at
100 Vac
DC Output 5V DC, 1A (5.0 Watts)
Electromagnetic Immunity and Emissions Declaration
and Guidance
The transmitter and receiver are intended for use in the electromagnetic environment
specified in the next table. The customer or the user of the transmitter should ensure
that it is used in such an environment.
Immunity Test
Transmitter Compliance
Level
Receiver Compliance
Level
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
Magnetic Field (50Hz)
IEC 61000-4-8
30 A/m
Electrical Fast
Transient/Burst
IEC 61000-4-4
N/A
± 2 kV for power supply
lines
Surge
IEC 61000-4-5
N/A
± 0.5 kV, ± 1 kV line(s) to
line(s)