91Appendix H: Professional Use Instructions 91
Dexcom G6
®
System • Using Your G6
4 Disinfect Transmitter
a. Prepare
Protect:
Wear clean gloves and goggles
Prepare Soak:
Put CaviCide
®
solution (Cavicide) in a container deep enough to
submerge the transmitter
b. Clean
Scrub:
Saturate clean cloth or wipe
with Cavicide
Wipe entire transmitter for
at least 3 minutes or until all
soil is removed
Focus on the nooks and
crannies – see green arrows
Soak:
Put transmitter in prepared
soak
Swirl it for 30 seconds
Then let it soak for another
2 ½ minutes
c. Rinse and Dry
Rinse:
Remove from soak and rinse transmitter under flowing tap water for
10 seconds
Dry:
Wipe transmitter dry with cloth
The transmitter was tested through 11 clean-disinfect-use cycles.