91Appendix H: Professional Use Instructions 91
Dexcom G6
®
System • Using Your G6
4 Disinfect Transmitter
a. Prepare
• Protect:
• Wear clean gloves and goggles
• Prepare Soak:
• Put CaviCide
®
solution (Cavicide) in a container deep enough to
submerge the transmitter
b. Clean
• Scrub:
• Saturate clean cloth or wipe
with Cavicide
• Wipe entire transmitter for
at least 3 minutes or until all
soil is removed
• Focus on the nooks and
crannies – see green arrows
• Soak:
• Put transmitter in prepared
soak
• Swirl it for 30 seconds
• Then let it soak for another
2 ½ minutes
c. Rinse and Dry
• Rinse:
• Remove from soak and rinse transmitter under flowing tap water for
10 seconds
• Dry:
• Wipe transmitter dry with cloth
The transmitter was tested through 11 clean-disinfect-use cycles.