90 Appendix H: Professional Use Instructions
Dexcom G6
®
System • Using Your G6
b. Clean
• Position Transmitter:
• Put transmitter, Dexcom side up, on clean surface or absorbent wipe
• Spray:
• Spray top of transmitter twice with Clorox, holding bottle 15–20
centimeters from transmitter
• Position Transmitter:
• Turn transmitter over, with metal dots up
• Spray:
• Spray top of transmitter twice with Clorox, holding bottle 15–20
centimeters from transmitter
• Scrub:
• For 30 seconds, scrub all sides of transmitter with a nylon brush
• Soak:
• Put transmitter in prepared soak for 1 minute
c. Rinse and Dry
• Rinse:
• Remove from soak and rinse transmitter under flowing tap water for
10 seconds
• Dry:
• Wipe transmitter dry with cloth
d. Inspect
• Verify there is no visible soil. If there is, reclean.