DIESTRO 103APV4R ELECTROLYTE ANALYZER
27 JS Medicina Electrónica
- The analyzer then automatically loads the StdA needed for rinsing and one point
calibration.
If the analyzer is not able to load Std. A will display the "Not Filled" error and
the measurement will not be completed.
See "Error Messages" section for more information
- When the measurement is complete, the analyzer displays the result on the screen, and
prints the ticket. The time that the results stays on screen is configurable.
If the Sodium result is lower than expected and you have a Sodium Glass
Electrode, perform a Sodium conditioning wash.
See section "Sodium Conditioner"
3.6 Loading patient data:
You can enter the data by hand, by pressing on each of the fields, or directly by scanning the
label of the tube. The data you can enter are:
• id number
• Name (letters and numbers)
• Sex
• Age (age range)
In case 10 minutes of inactivity have passed since the last operation the
equipment will automatically add a rinse before the measurement.
BIOHAZARD. Samples, capillaries and adapters are potentially infectious.
Handle with gloves.