DIESTRO 103APV4R ELECTROLYTE ANALYZER
56 JS Medicina Electrónica
13 - NON-DETECTABLE SAMPLES
1. OVERVIEW
It may happen that a sample (Serum, Whole Blood or Urine) with low conductivity or insufficient
volume is not detected giving an error in the analyzer (Error: Not Filled).
The operator can position the sample manually to perform the measurement.
2. MANUAL POSITIONING
2.1 Faced with a "Not Filled" error, the analyzer displays the following window:
If you press "CANCEL", the analyzer will run a rinse and the sample will be lost. After rinsing,
the analyzer will be ready to measure again
If you want to position manually:
2.2 Open the front of the analyzer to get a view of the electrode chamber.