DIESTRO 103APV4R ELECTROLYTE ANALYZER
33 JS Medicina Electrónica
3. TWO POINT CALIBRATION OF IONS AND PH
3.1 Automatic Calibration
It is performed when the analyzer is turned on, and every 8 hours.
The calibration frequency can be modified in the calibration options menu.
Route: "HOME" screen → Calibration Options → Calibration Frequency
3.2 On-demand calibration
Calibration can be performed on demand. To do this, navigate to the Calibration screen:
Route: "HOME" screen → → Calibration
Then press " Electrolyte Calibration" to begin calibration.
If the analyzer is not able to load Std. A or Std.B will show a "Not Filled" error
and calibration will not be performed. View maintenance errors
4. TCO2 CALIBRATION
The calibration of TCO2 is carried out in the factory and in special situations (maintenance or
service of the TCO2 Module). Check with your dealer when performing this calibration.
To do this, navigate to the calibration screen and press "TCO2 Calibration"