DIESTRO 103APV4R ELECTROLYTE ANALYZER
84 JS Medicina Electrónica
4. SAMPLE mV MEASUREMENT
Measure the voltages corresponding to the electrodes installed in the equipment.
The mV shown (mV difference between the electrode in question and the reference electrode),
are proportional to the concentration of each ION (Interpretation of the results obtained).
The result obtained for the Std. A and the Std. B, allow to verify the gain of each electrode in
mV.
See section:
"TECHNICAL SPECIFICATIONS" → "Electrode Gain Range"