DIESTRO 103APV4R ELECTROLYTE ANALYZER
36 JS Medicina Electrónica
6. PRINTING THE CALIBRATION RESULT
The printed output includes the following information:
=======================
Calibration Report
#125 Calibration ID
04/04/2022 10:08:18 Calibration Date & Time
JS Medicina Electronica Institution Name
103APV4R S/N: 1 Analyzer Model and Serial Number
UI: 1.0 uISE: 1.57 UI & Firmware Versions
-----------------------
Pack LOT: 8 Pack Lot Number
Expires : 2023/09/01 Expiration Date
Std. A : 87% B: 34% A & B Reagents Level
-----------------------
Na = Calibrated Ion Name & Calibration Status
Slope : 71.2 Slope
Balance : +24.84 mV Balance
-----------------------
K = Calibrated
Slope : 63.6
Balance : -4.98 mV
-----------------------
Cl = Calibrated
Slope : 58.2
Balance : +21.77 mV
-----------------------
Ca = Calibrated
Slope : 29.9
Balance : +7.97 mV
-----------------------
Li = Calibrated
Slope : 46.9
Balance : -12.74 mV
-----------------------
pH = Calibrated
Slope : 52.0
Balance : +76.41 mV
Temp. : 20.2 °C
=======================