DIESTRO 103APV4R ELECTROLYTE ANALYZER
34 JS Medicina Electrónica
Before calibrating, the analyzer will wash the TCO2 module and display the following screen
Wait until the rinse is finished. The Analyzer will then ask you to lift the lever and place the
calibration solution. Follow the on-screen instructions.
Use only the TCO2 calibration solution provided
by the manufacturer
5. CALIBRATION RESULT
Once the Calibration is completed, the Analyzer will print a ticket with the result. If the
calibration was executed manually, the results will be displayed on the screen.
The analyzer will display the status, gain, and balance. If any result is not valid, the electrode
will say "Uncalibrated", turn red and it will not be possible to measure this parameter.