DIESTRO 103APV4R ELECTROLYTE ANALYZER
71 JS Medicina Electrónica
Failure of
electronic
components
Technical Service.
C
The balance
during 1-point
calibration
differs from the
last 2-point
calibration.
External
interference
during
measurement.
Bubbles in the
liquid circuit.
Electrode failure.
Failure in the
measurement
channel.
Poor grounding.
Supply voltage
with interference.
The result is
not valid.
Repeat
measurement or
calibration
Check for blockages
or leaks in pipes
and electrodes.
Check valves and
peristaltic pipe.
Change electrodes.
Investigate possible
interference from
other equipment or
appliances.
Contact technical
support.
L
The balance
during 1-point
calibration
differs from the
last 1-point
calibration.
Q
Sharp
differences
during the
measurement
process.
In the
Sharp
differences
during sample
stabilization.
Measures low
sodium
Dirty sodium
electrode
Low sodium
electrolyte
measurement
values
Pass sodium
conditioner
18 MAINTENANCE
The analyzer was designed to require minimal and easy maintenance:
1. DAILY MAINTENANCE
1.1 Analyzer decontamination
BIOHAZARD. Samples, capillaries and adapters are potentially infectious.
Handle with gloves.
1.1.1 Keep the workbench and analyser surfaces in hygienic conditions.

Table of Contents