DIESTRO 103APV4R ELECTROLYTE ANALYZER
70 JS Medicina Electrónica
!
inverted.
changed or
contaminated.
measured.
Pack.
Purge and
calibrate.
Change for a new
Pack.
Not Full
The sample or
calibrator
solution could
not be loaded
normally.
Insufficient
sample.
Sample with
clots, fibrins, or
bubbles.
Operator error.
Electrodes or
plugged pipe
circuit.
Peristaltic pipe or
faulty valves.
The ongoing
operation will
be aborted.
Verify the sample
Verify the charging
process.
Check for
blockages in pipes,
electrodes, or
peristaltic pipe.
Check valves.
Change the
peristaltic pipe.
Change the capped
electrode
Contact technical
support.
Not
empty
The measuring
chamber
cannot be
emptied.
Electrodes or
plugged pipe
circuit.
Peristaltic pipe or
faulty valves.
The ongoing
operation is
aborted.
Missing
Pack
It does not
recognize the
uChip.
uChip not
connected.
Failure of
electronic
components
It does not
calibrate.
It doesn't
measure.
No lava.
Connect a uChip.
Turn off and back
on. (*)
Contact the
Technical Service.
Error
Description
Possible Cause
Effect
Action
Expired
Pack
Pack Expired.
Pack out of date.
Incorrect date on
the analyzer.
Failure of
electronic
components
Prints the
legend "Pack
Expired" on
the screen and
on the printer
when
calibrating.
Check the date of
the Pack.
Turn off and back
on. (*)
Set the analyzer
date.
Change for a new
Pack.
Contact the
Technical Service.
Pack
Sold Out
The Pack is
empty.
In the uChip
some of the
calibration
solutions are
exhausted.
It does not
calibrate.
It doesn't
measure.
No lava.
Change for a new
Pack.
Turn off and back
on.
Contact the