DIESTRO 103APV4R ELECTROLYTE ANALYZER
41 JS Medicina Electrónica
● Calibrations performed.
● Wets of Std A.
● Samples in Batch mode.
● Samples in random mode.
● Normal washes.
● Wets of Std B.
● "Not empty" errors
● "Not filled" errors
This allows you to verify the use of the analyzer.
To access this information click on the pack status indicator, from any menu.
11. STANDBY FUNCTION
After 10 minutes of inactivity, the analyzer automatically goes into standby.
While in this mode, automatic calibration and washing operations are suspended.
In this mode two types of operations will be executed:
- A&B wetting: performed every 8 hours after the last calibration
- Wetting A: performed every 3 hours after the last calibration.
It is recommended to keep the analyzer on 24 hours a day, 365 days a year.