DIESTRO 103APV4R ELECTROLYTE ANALYZER
44 JS Medicina Electrónica
9 PURGES
To access this menu, navigate to:
Route: "HOME" screen → → Rinse
1. PURGE STDS
During this process, the pack and equipment pipes are filled with both calibration solutions.
2. PURGE LACTIC
In equipment equipped with a module for measuring TCO2, the lactic acid pipe circuit must be
purged when changing the Lactic Acid bottle.
If the equipment has not been used for a long time you must also perform a purge.
After the purge is complete, a TCO2 wash is carried out automatically.
3. AUTOMATIC PURGE
It is performed automatically when a new pack is installed in the analyzer,
The analyzer will first purge the reagent from the stdB bottle and then the reagent from the stdA
bottle.
When the process is finished, a calibration will automatically begin.
If for some reason the pipes of the Pack are disconnected, it is necessary to
perform a purge manually.
Otherwise, calibration, measurement and/or rinsing will fail