DIESTRO 103APV4R ELECTROLYTE ANALYZER
30 JS Medicina Electrónica
Slide the display until you see the pH electrode, then select between 37°C and room
temperature. If you select 37°C, the analyzer will calculate the corrected pH at that temperature.
For more details, see section:
"SERUM/PLASMA/WHOLE BLOOD MEASUREMENT"
5.5. Quality control:
Samples measured from the quality control menu are always reported at 37°C. The Quality
Control of Corrected Ca is carried out in the same way as the other Ions
For more details, see section:
"QUALITY CONTROL"
6 - URINE MEASUREMENT
1. OVERVIEW
Ensure the correct functioning of the analyzer by servicing it correctly and
performing a quality control.
See the sections: "QUALITY CONTROL" and "MAINTENANCE"
Always dilute urine samples. The default dilution is 1 part urine with 4 parts
diluent (1:5).
The analyzer only measures Na, K, and Cl in urine samples.
Use Urine Diluting Solution ISE REF IN 0300