DIESTRO 103APV4R ELECTROLYTE ANALYZER
72 JS Medicina Electrónica
1.1.2 Clean all outer surfaces of the analyzer with a cloth gently moistened with 1:10
solution of sodium hypochlorite.
1.1.3 Decontaminate the Fill Port with ISE Cleaning Solution Solution ISE REF IN 0400.
1.2 ISE Cleaning
To protect the analyzer from possible contamination and from obstructions and plugs, perform
ISE Cleaning daily.
It is recommended to do the ISE Cleaning at the end of the working day, to remove possible
residues from the liquid circuit.
If ISE Cleaning is not performed daily, the analyzer will not allow to continue
working until one is completed.
See sections:
"ISE CLEANING AND RINSING".
2. WEEKLY MAINTENANCE
2.1 Sodium conditioning solution
Once a week perform an intensive wash with the sodium conditioning solution.
See section:
"SODIUM CONDITIONER".
2.2 Internal cleaning of the equipment
Clean the surface that says "keep clean" and any splashes of biological substances with a cloth
moistened in a 1:10 dilution of sodium hypochlorite.
3. OTHER MAINTENANCE AND REPLACEMENT OF SPARE PARTS OR
COMPONENTS
BUY ONLY ORIGINAL SPARE PARTS.
The frequencies indicated are those recommended, but they are only preventive. These
frequencies can be modified depending on the number of samples processed and according to
the needs.