DIESTRO 103APV4R ELECTROLYTE ANALYZER
73 JS Medicina Electrónica
The following table provides instructions, recommended frequencies, and who can make the
changes:
Precaution. Potentially infectious elements, discard according to the laws of
your country for the treatment of pathological waste. Handle with gloves.
Spare Part /
Component
Recommended
Frequency
Authorized Person
Peristaltic pump
head
Every 6 months
Trained operator
Distributor
Technical Service
Maker
Electrodes
According to the
need, they have
an estimated
shelf life of one
year
Trained operator
Distributor
Technical Service
Maker
Capillary
sampling
Every 6 months
Trained operator
Distributor
Technical Service
Maker
Standard and
valve pipes
Every 1 year
Distributor
Technical Service
Maker
Stack
According to the
need
Distributor
Technical Service
Maker
Fill Port
According to the
need
Trained operator
Distributor
Technical Service
Maker
Sampling cleaner
Simultaneous
with the
DIESTRO Pack
Trained operator
Distributor
Technical Service
Maker
NiMh Battery
(optional)
Every 3 years
Trained operator
Distributor
Technical Service
Maker