DIESTRO 103APV4R ELECTROLYTE ANALYZER
57 JS Medicina Electrónica
2.3 Using the Left and Right buttons, position the sample so that it is in contact with the
reference electrode and the electrodes for which it reaches volume.
2.4 Once the sample is positioned correctly and without bubbles, lower the lever and press
"Measure".
2.5 The measurement will begin and the Analyzer will continue normally
The operator is responsible for positioning the sample correctly. Position it
from the reference electrode covering as many electrodes as possible with the
amount of sample it counts. Remember that the sample has to be in contact
with the reference electrode. Beware of electrodes that are not in contact with
the sample. The results of such electrodes will be incorrect. It is the
responsibility of the operator to discard the results of such electrodes.