DIESTRO 103APV4R ELECTROLYTE ANALYZER
76 JS Medicina Electrónica
The needle and the used Sampling Cleaner may contain potentially infectious
residues, exercise caution, always wear gloves, do not splash.
Discard the cloth or towel and the sample cleaner used following the
clarifications in the section "DISPOSAL OF SUPPLIES".
6.4 Open the transparent packaging and remove the safety seal from the new Sample
Taking Cleaner as shown in the figure.
6.5 Present the DIESTRO cleaning module on the front of the position guides, and slide
smoothly until you stop against the bottom.
Make sure the Cleaning Module is in the correct position, labeled "Front/Front"
towards the operator.
6.6 Introduce the capillary sample into the analyzer.
6.7 Close the front of the analyzer, turn it on and continue using the equipment normally.