SPEC 605: SELECTION OF MANUAL SCALE START
0: AUTO START
1: MANUAL START
SPEC 606: SELECTION OF ZERO LAMP LIGHTING POSITION
0: LIGHT ON AT GROSS ZERO ± 1 / 4 DIGIT
1: LIGHT ON AT NET ZERO ±1/4 DIGIT
SPEC 607: SELECTION OF DECIMAL POINT POSITION FOR WEIGHT
0: NONE
1: 0.0
2: 0.00
3: 0.000
SPEC 608: SELECTION OF ZERO POINT SETTING RANGES WHEN A/C DISPLAY SWITCH ON
0: ±10% OF CAPACITY (CALIBRATED ZERO POINT ± 6000IR)
1: ±5% OF CAPACITY (CALIBRATED ZERO POINT ± 3000IR)
2: ±2% OF CAPACITY (CALIBRATED ZERO POINT ± 1200IR)
3: ±0.6% OF CAPACITY (CALIBRATED ZERO POINT ± 360IR)
SPEC 609: SELECTION OF DECIMAL POINT POSITION FOR UNIT PRICE & TOTAL PRICE
0: NONE
1: 0.0
2: 0.00
3: 0.000
SPEC 610: SELECTION OF UNIT PRICE DECIMAL POINT RIGHT SHIFT
0: NO RIGHT SHIFT
1: RIGHT SHIFT
SPEC 611: SELECTION OF UNIT PRICE DECIMAL POINT LEFT SHIFT
0: NO LEFT SHIFT
1: LEFT SHIFT
SPEC 612: SELECTION OF SWITCHING OF PRICE BASE
0: INHIBIT
1: 100g / 1 KG
2: 500g / 1 KG
SPEC 613: SELECTION OF EXIT FROM ACCUMULATION MODE AFTER TIME OUT
0: INHIBIT
1: 15 SECS
2: 5 SECS
3: 10 SECS
SPEC 614: SELECTION OF ZERO SETTING RANGES WHEN WEIGHT RESET SWITCH IS ON OR ZERO
TRACKING
0: ±10% OF CAPACITY (POWER ON START POINT ± 6000IR)
1: ±5% OF CAPACITY (POWER ON START POINT ± 3000IR)
2: ±2% OF CAPACITY (POWER ON START POINT ± 1200IR)
3: ±0.6% OF CAPACITY (POWER ON START POINT ± 360IR)
SPEC 615: SELECTION OF CAPACITY
0: 3Kg
1: 6Kg
2: 15Kg
3: 30Kg
4: 30LB
SPEC 616: SELECTION OF WEIGHT DUAL / SINGLE RANGE
0: SINGLE
1: DUAL
2: 1 / 7500 OR 1 / 6000