136 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Tests
7.5.6.3 Sample line
Prerequisite: The device is equipped with the "integrated patient-gas measurement
module".
Check that the sample line is correctly connected.
7.5.6.4 Water trap
Prerequisite: The device is equipped with the "integrated patient-gas measurement
module".
1. Check the water level in the water trap.
2. Check the period of use of the water trap. Replace the water trap when
necessary.
With Infinity ID function:
Water traps of type Infinity ID WaterLock 2 will automatically be detected and
the replacement date will automatically be set.
Without Infinity ID function:
Water traps will not be detected.
Update the replacement date manually: Touch the Reset button after a new
water trap has been installed.