Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 179
Operation
Manual setting
In the Alarms dialog, the alarm limits can be set either manually or automatically.
1. Open the Alarms dialog.
2. Open the Limits tab (1).
3. Set the upper alarm limit (2).
4. Set the lower alarm limit (3).
Automatic setting
Alarm limits can be automatically adapted to current measured values or set values.
1. Open the Alarms > Limits dialog.
2. To adjust the alarm limits for one individual parameter, touch and confirm one of
the Autoset buttons (5).
To adjust the alarm limits for all parameters, touch and confirm the Autoset all
button (4).
The function for adjusting all alarm limits can be called up directly with the Autoset
limits button in the main menu bar.
Only use the automatic adjustment when measured values and set values are
stable.
The lower alarm limit for the xMAC level is also adjusted during automatic setting,
see page 180.
For the configuration and algorithm, see page 208.
8.10.2 Activating and deactivating CO2 alarms
Prerequisite: The device is equipped with the "integrated patient-gas measurement
module".
CO
2 monitoring (alarms for inCO2, etCO2, and CO2 apnea) can be activated or
deactivated.
1. Open the Alarms dialog.
35935
45
2
3
1
Alarms

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals