EasyManuals Logo

Dräger Atlan A100 User Manual

Dräger Atlan A100
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
180 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Operation
2. Touch the Settings tab (1).
3. For CO
2 alarms, touch the button (2):
On: Alarms are activated.
Off: Alarms are deactivated.
Or
Use the CO
2 alarms off button in the main menu bar to deactivate or activate
the alarms.
This button is only visible in the following ventilation modes:
Manual / Spontaneous
External fresh-gas outlet
–Pause
1)
Monitoring
2)
The alarm system is immediately activated when the CO2 monitoring is activated.
Deactivation is indicated in the header bar and in the parameter field by the
symbol.
8.10.3 Automatic xMAC monitoring
Prerequisite: The device is equipped with the "integrated patient-gas measurement
module".
The xMAC monitoring is automatically activated as soon as the following conditions
are met:
Anesthetic gas is administered.
The inspiratory xMAC value is greater than the expiratory xMAC value.
The expiratory xMAC value reaches approximately 0.3.
35936
1) Pause must be activated in the system setup.
2) Only for devices with integrated PGM. Pause must be deactivated in the system setup.
1
2
Alarms

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals