Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n 193
Alarms
10.3.2 Silencing the alarm tone
The alarm tone can be silenced for a maximum of 2 minutes.
Press the alarm silence key (1) below the screen, see page 21.
The symbol and the remaining time for the silenced alarm tone are displayed in
the header bar.
If the cause of the alarm persists, the alarm tone starts again immediately after the
alarm silence ends.
During the alarm silence, only new alarms with a higher alarm priority or a higher
internal priority number compared with the silenced alarm are acoustically signaled.
For further information see: "Alarm – Cause – Remedy", page 245.
10.3.2.1 Reactivating the alarm tone
Press the alarm silence key again.
10.3.3 Downgrading and resetting alarm messages
Some alarms can be downgraded to low priority or they can be cleared completely.
The relevant alarms can be recognized in the table "Alarm – Cause – Remedy" on
page 245 by the following remedial messages:
36004
1
Remedial message Effect
Use "ALARM RESET" to downgrade
alarm priority.
Alarm priority is changed to low.
Use "ALARM RESET" to reset alarm. Alarm is cleared.

Table of Contents

Other manuals for Dräger Atlan A100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A100 and is the answer not in the manual?

Dräger Atlan A100 Specifications

General IconGeneral
BrandDräger
ModelAtlan A100
CategoryMedical Equipment
LanguageEnglish

Related product manuals