250 Instructions for use | Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Troubleshooting
!! 180 Battery charge low The battery charge is low
and the mains power supply
is not available.
Make sure that the mains
power supply is correctly
connected.
The breathing system
warmer has been switched
off. Check the breathing cir-
cuit for condensate. Increase
the fresh-gas flow if neces-
sary.
Use "ALARM RESET" to
downgrade alarm priority.
!!! 30 Battery charge very low The battery charge is critical
and the mains power supply
is not available. The device
may shut down in the next
minutes.
Restore mains power supply.
The breathing system
warmer has been switched
off. Check the breathing cir-
cuit for condensate. Increase
the fresh-gas flow if neces-
sary.
Once the battery has been
depleted, ventilate the patient
manually.
!! 170 Battery failure The battery is faulty. If the
mains power supply fails, the
device will switch off immedi-
ately.
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
! 100 Battery temperature high The battery temperature is
high. Charging of the battery
has been suspended to pro-
tect it from damage.
Ensure that the system is
connected to the mains
power supply.
Check the ambient tempera-
ture.
Use "ALARM RESET" to
reset alarm.
! 100 Breathing circuit expired Accessory has been used
too long.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
reset alarm.
Priority Alarm Cause Remedy