20 Technical manual | X-am® 3500/8000
Operation
2. Select and confirm Switch off.
The gas detector may only be switched off without a prior sign-on if the Switch-
off allowed feature is activated using the Dräger CC-Vision PC software. Factory
setting: activated
When the gas detector is placed in the charging cradle, it switches off
automatically.
If the gas detector remains off for longer than 21 days and is not charged, deep
sleep mode is activated. In deep sleep mode, the gas detector can no longer be
switched on using the Dräger CC-Vision PC software or the Dräger X-dock. In this
case, the gas detector must be switched on manually.
4.5 Logging a user in or out
The gas detector has four configurable user levels. The user levels can be set using
the Dräger CC-Vision PC software. User level 0 means that the user is not signed
on. User levels 1 to 3 require a password for sign-on.
The following passwords are created by default:
Dräger recommends changing the predefined passwords after initial
commissioning.
Default setting:
User level 1: 0001
User level 2: 0002
User level 3: 0003
Function User level
0
Bump test
---
Fresh air adjustment
--
Clearance measurement
1)
1) X-am 8000 only
--
Leak search
1)
---
Benzene/pre-tube measurement
1)
---
Settings menu
---
Maintenance menu
---
Change measurement gas
1)2)
2) The data stored in the statistics counters are lost when the measurement gas is changed. The
Dräger GasVision PC software can be used to manually view the data via the data logger at a
later point in time. Automatic notifications via the X-dock Manager may only be available to a
limited extent.
---
Bluetooth
®1)
--