56 Technical manual | X-am® 3500/8000
Maintenance
6.9.4 RFID transponder
Requirements:
The gas detector is open.
1. Loosen 3 screws (Torx T8) on the options board.
2. Rotate the gas detector and carefully lift it out of the front housing cradle.
3. Remove the options board cover.
4. Detach the RFID transponder (1) from the bracket in the options board cover
(top end).
5. Insert the new RFID transponder in the mount.
6. Read the ID of the RFID transponder and save it to the device using the Dräger
CC-Vision PC software.
6.9.5 Front housing cradle with diaphragms
Requirements:
The gas detector is open.
1. Loosen 3 screws (Torx T8) on the options board.
2. Rotate the gas detector and carefully lift it out of the front housing cradle.
3. Exchange the front housing cradle.
The sensor and pump diaphragms can only be exchange with the upper cradle.
6.9.6 Vibration motor
Requirements:
The gas detector is open.
1. Loosen 3 screws (Torx T8) on the options board.
2. Rotate the gas detector and carefully lift it out of the front housing cradle.
36329
1

Table of Contents

Other manuals for Dräger HFG 000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger HFG 000 Series and is the answer not in the manual?

Dräger HFG 000 Series Specifications

General IconGeneral
BrandDräger
ModelHFG 000 Series
CategoryGas Detectors
LanguageEnglish

Related product manuals