Technical manual | X-am® 3500/8000 65
Configuration
7.2 Configuring the gas detector
When the capture range is deactivated, the data memory fills up with log files
significantly faster, making is necessary to read it out more frequently.
When a configuration is transmitted to the gas detector with the CC-Vision PC
software, existing TWA and STEL evaluations are reset.
7.2.1 Configuring the gas detector with the Dräger X-dock and reading
the data memory
To configure a gas detector with the X-dock maintenance station and read the data
memory, insert the gas detector in the X-am 8000 module of the maintenance
station.
Refer to the Data dongle X-dock Instruction for Use for information on how to
configure the gas detector or read the data memory with the maintenance stations.
Methyl tert-butyl
ether (MTBE)
1634-04-
4
MTBE 0.87 50 100 1.06 4 8
n-Nonane 111-84-2 Nona 1.55 50 100 2.1 10 20
n-octane 111-65-9 Octa 1.93 50 100 2.14 10 20
Alpha-pinene 2437-95-
8
aPin 0.38 50 100 0.48 2 4
Styrene 100-42-5 Styr 0.44 50 100 0.82 4 8
Toluene 108-88-3 Tolu 0.56 50 100 0.98 4 8
Trichloroethylene 79-01-6 TCE 0.53 50 100 0.95 4 8
Vinyl chloride 75-01-4 VC 1.97 50 100 2.18 10 20
o-xylene 95-47-6 Xyol 0.55 50 100 0.79 4 8
Diesel 68476-
34-6
Desl 1 50 100 1 4 8
Petrol 8006-61-
9
Gaso 1.21 50 100 1.21 4 8
Jet fuel 8008-20-
6
JetF 1 50 100 1 4 8
1.3-butadiene 106-99-0 BTD1 0.69 50 100 0.69 2 4
1-butanol 71-36-3 BuOH 4.93 50 100 5.27 10 20
Isooctane 540-84-1 iOct 1.19 50 100 1.45 4 8
1.1-dichloroethene 75-35-4 DCE 0.85 50 100 0.76 4 8
Tetrahydrofuran 109-99-9 THF 2 50 100 1.77 10 20
Dimethylformamide 68-12-2 DMF 0.95 50 100
---
2)
---
2)
---
2)
Cyclohexanone 108-94-1 CyHo 0.97 50 100 1.08 4 8
PID - PID 1 50 100 1 4 8
1) RF: Response factor at 20 °C relative to isobutylene (CAS no.: 115-11-7).
2) The measuring capability of the sensor type is not sufficient for this substance.
PID HC PID LC ppb
Substance CAS no.
RF
1)
A1 [ppm] A2 [ppm]
RF
1)
A1 [ppm] A2 [ppm]