Technical manual | X-am
®
2800 / X-am
®
5800 11
Operation
1. Hold down the button for approx. 3 s. The display shows the following
screens one after the other.
Countdown
The switch-on sequence and the warm-up phase of the sensors start.
Display test
Start screen
Firmware version
Alarm element test (LEDs, alarm signal and vibration alarm)
Customer-specific information screen (optional and can be configured with
the Dräger CC Vision PC software)
Measuring channel, sensor type, alarm thresholds, STEL, TWA (if
configured) and LEL factor (if available)
Any expired bump test or calibration interval as well as early warnings (if
configured)
Measured value display
4.3.2 Switching off the gas detector
When the gas detector is placed in the charging cradle, it switches off
automatically (only when the automatic distress signal is deactivated).
1. Hold down and simultaneously until the displayed countdown has
finished.
The visual, acoustic and vibration alarms activate briefly.
The gas detector is switched off.
4.4 Preparations for operation
WARNING
Serious damage to health
An incorrect calibration can lead to incorrect measured values, which may result in
serious damage to health.
Before performing safety-relevant measurements, check the calibration by
means of a bump test, calibrate as necessary, and check all alarm elements. If
national regulations exist, the bump test must be performed in accordance with
these regulations.
The gas detector should be worn for personal monitoring in the proximity of the
breathing zone.
Dräger recommend using the clip to attach the gas detector to clothing either on
your collar or breast pocket. If gases are expected in a closed space that are
(much) heavier than air (e.g. CO
2), then the gas detector can be worn in a lower
position, e.g. on your belt or at your hip.
The gas detector has magnetic switches. Do not put any magnets in the
immediate proximity of the gas detector (e.g. name badges with magnetic holders).
This could trigger features (e.g. bump test) on the gas detector.
OK
OK

Table of Contents

Other manuals for Dräger MQG 02 0 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger MQG 02 0 Series and is the answer not in the manual?

Dräger MQG 02 0 Series Specifications

General IconGeneral
BrandDräger
ModelMQG 02 0 Series
CategoryGas Detectors
LanguageEnglish

Related product manuals