Technical manual | X-am
®
2800 / X-am
®
5800 21
Troubleshooting
165 Error sent from Bluetooth control. Switch the gas detector off and
then on again / Contact Dräger-
Service.
175 Bluetooth connection interrupted. Reconnect the device.
176 Subsystem software version too
old.
Carry out a software update.
182 Device temperature too high. Operate gas detector within per-
mitted temperature range.
183 Device temperature too low. Operate gas detector within per-
mitted temperature range.
185 Battery pre-alarm Charge or replace battery.
x50 Lamp intensity too low. Clean the lamp or replace the sen-
sor.
x51 Sensor warming up (warm-up 1).
Followed by measurement
standby, bump test standby
Wait until warm-up time is com-
plete.
x52 Sensor warming up (warm-up 2).
Followed by calibration standby
Wait until warm-up time is com-
plete.
x53 Value negative. Perform a fresh air calibration.
x54 Temperature too high. Operate gas detector within per-
mitted temperature range.
x55 Temperature too low. Operate gas detector within per-
mitted temperature range.
x56 Calibration interval close to expiry. Carry out the span calibration.
x69 Bump test expired. Carry out bump test.
x70 Bump test interval expires soon. Carry out bump test.
x71 Span calibration close to expiry
(CatEx thermal conduction range).
Carry out the span calibration.
x72 Sensor switched off due to over-
gassing.
Restart the gas detector.
x73 Calibration expired
(CatEx thermal conduction range).
Carry out the span calibration.
x77 Calibration interval has expired. Carry out the span calibration.
x78 Sensor software version is too old. Update the sensor.
x80 Sensor life time has expired. Replace the sensor as soon as
possible.
x82 Deviation after zero calibration of
the IR sensor.
Carry out span calibration for IR
sensor.
Code Cause Remedy