Technical manual | X-am
®
2800 / X-am
®
5800 37
Maintenance
c. Have the sensor type and code (marked on the new sensor) as well as the
sensor position at hand for the new sensor inauguration.
d. Carefully insert the new sensor.
e. Carefully plug the flexstrip into the bushing on the options board.
f. Close the sensor damper and make sure that it is positioned correctly.
8. Reattach the front and rear housing cradles and makes sure that the seals and
sensor dampers are in the correct positions.
9. Screw on the rear housing cradle with 4 screws (35 ±5 Ncm).
Dräger recommends first tightening the two lower housing screws and then
the two upper ones.
10.Follow the instructions in the sensor exchange wizard in the CC-Vision PC
software.
Next steps:
After every sensor exchange, the fresh air calibration/ zero calibration must first
be carried out followed by the span calibration (mixed-gas or single-gas
calibration).
For the XXS O
2, Dräger recommends testing the alarm feature with a suitable
test gas after the fresh air calibration.
51588